Recall Z-2957-2011 ยท FMI Needle, Hypodermic, Single Lumen

Greiner Bio-One North America, Inc.

HOLDEX Single-Use Holder PP closed pack. Greiner Bio-One Austria Gmbh. A sterile single-use disposable holder with an off-center female luer at the top of the holder and epoxy bonded stainless steel needle on the underside of the holder, recommended for use with commercially available butterfly and luer needles in routine venipuncture procedures.

DateAug 8, 2011 (initiated Jun 29, 2011)
StatusTerminated
ReasonAt removal of the tube from the holder, the needle may dislodge and blood spill can occur.
Root causeOther
ActionGreiner Bio-One North America, Inc. sent a "Recall HOLDEX Single Use Holder PP" letter dated June 29, 2011 to all affected customers. The letter instructed customers stop distributing the product immediately. A Product Disposition form was provided to be completed and returned to Greiner Bio-One at which time upon receipt, arrangements for product return to Greiner Bio-One will be made for replacement or credit. They were requested to provide a list of their customers to Greiner Bio-One that received the affected item. For additional information call Greiner Bio-One at (888) 266-3883.
Quantity142,400 pieces
DistributionNationwide Distribution including the states of CA, GA, IL, KS, MA, IL, PA and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980768VACUETTE HOLDEX