Recall Z-2970-2011 ยท LZS Excimer Laser System
AMO Manufacturing USA, LLC
STAR S4 IR Excimer Laser System; Software Version 5.16 Mfg by : AMO Manufacturing USA, LLC 510 Cottonwood Drive Milpitas, CA 95035 Iris registration is designed to compensate for pupil centroid shift and cycloration, both of Which may occur because of patient characteristics and the different environments at the time of measurement and at the time of treatment.
| Date | Aug 9, 2011 (initiated Jul 18, 2011) |
| Status | Terminated |
| Reason | Reports of lower success rates with Iris Registration (IR) image capture with the STAR S4 IR Excimer Laser System v 5.3 that with previous version. |
| Root cause | Software design |
| Action | Abbott Medical Optics (AMO) sent an "ADVISORY NOTICE" dated July 18, 2011 to all customers using the AMO WAVESCAN Wavefront System. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to create a separate patient file for each eye on the WAVESCAN System (STAR S4 IR Excimer Laser Systems,software versions 5.00-5.30) according to the directions in attachment A. Additionally, customers are to complete and return an attached form to AMO via fax at the regional fax number. Contact an AMO representative for questions regarding t |
| Quantity | 673 total - v5.00 -5.22 |
| Distribution | Worldwide Distribution-USA (nationwide) including Puerto Rico and countries of Africa, Asian Pacific, Canada, Europe, Japan, Latin America, and the Middle East. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.