Recall Z-0039-2012 ยท IDO Microtome, Rotary

Richard-Allan Scientific Company

Thermo Scientific Microm: Rotary Microtome: REF HM355S-3 Used for slicing tissue blocks.

DateOct 12, 2011 (initiated Aug 12, 2011)
StatusTerminated
ReasonThe firm has received two complaints of finger lacerations in operators of this device.
Root causeDevice Design
ActionThermoFisher Scientific sent "URGENT MEDICAL DEVICE CORRECTION, RE: Thermo Scientific HM355S-3 User Safety Advisory Notice" to all affected customers. The letter states the manufacturer identified a need to enhance the safety instructions and the description of the safety devices in the user manual. It mentions the HM3555S-3 User Manual has been modified to include more specific safe work practices and a new Quick -Reference Card summarizing the manual's safety enhancements and that the changes have been included on a CD that is enclosed with the letter. The firm requests confirmation of r
Quantity786
DistributionWorldwide Distribution: Nationwide distribution; including the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.