Recall Z-3265-2011 ยท MKJ Automated External Defibrillators (Non-Wearable)

Cardiac Science Corporation

Powerheart Automated External Defibrillator (AED), G3 is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of the ECG rhythm. If a shockable ventricular tachyarrhythmia recurs, the device will charge a

DateSep 19, 2011 (initiated Jun 30, 2011)
StatusTerminated
ReasonCardiac Science Corporation (CSC) has identified two errors, 0x48 and 0x99, that have repeatedly occurred during a device self-test in the G3 installed base over the last 6 years. These self test errors cause the device to enter into a service required state. CSC remind owners/operators of the Powerheart AED devices to verify the "Rescue Readiness" of their equipment and follow maintenance pro
Root causeDevice Design
ActionCardiac Science began mailing out the recall notification letter on August 1, 2011, to all their US consignees via USPS first class mail. Cardiac Science's distribution partners will receive a notification letter and be provided with a template letter to send to their customers directly. CSC remind owners/operators of the Powerheart AED devices to check the status of of the AED every day to ensure it is Rescue Ready, and perform the regularly schedule maintenance as outlined in the Powerheart AED "Operator and Service Manual" provided with every AED. Customers can call Cardiac Science at
Quantity195,739 units in US
DistributionWorldwide Distribution - US (nationwide) This action applies only to domestic
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031987POWERHEART AED G3, MODEL 9300
Cardiac Science, Jul 30, 2003
K040438POWERHEART AED G3A
Cardiac Science, Inc., Jul 1, 2004
K052161POWERHEART AED G3 (MODEL 9390E), POWERHEART AED G3 AUTOMATIC (MODEL 9390A)
Cardiac Science, Inc., Oct 21, 2005