Recall Z-2976-2011 ยท IZI System, X-Ray, Angiographic

Philips Healthcare Inc.

Allura Xper Systems Models: Allura Xper FD10 (F/C)-Biplane Fixed Allura Xper FD10/10 (incl. OR table) Allura Xper FD20- High resolution fixed-x ray Allura Xper FD20/10 Allura Xper FD20/20 Allura CV20 The Allura Xper F010, FD10/10, FD20, FD20/10 and FD20/20 Use From Labeling): is intended for use in cardiovascular and vascular X-ray imaging applications. including diagnostic,interventional proced

DateAug 10, 2011 (initiated Jul 8, 2011)
StatusTerminated
ReasonSystem freezes from 20 seconds up to one minute. No fluoroscopy exposure and no geometry movements during this time period.
Root causeSoftware design
ActionPhilips Health Care North America sent a FIELD SAFETY CORRECTION ACTION letter dated July 4, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. This FCD will be identified as FCD72200200 and will be free of charge. This FCD consists a CD with the correct firmware and a installation procedure to implement the FCD correctly. Please contact your local Philips representative. For further follow up please call (978 ) 687-1501.
Quantity1423 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031333PHILIPS INTEGRIS ALLURA FLAT DETECTOR RELEASE 1.2
Philips Medical Systems North America, Inc., May 13, 2003
K033737ALLURA XPER FD20
Philips Medical Systems North America, Inc., Dec 9, 2003
K041949ALLURA XPER FD10
Philips Medical Systems North America Co., Jul 30, 2004