Recall Z-2865-2011 ยท FPO Stretcher, Wheeled

Stryker Medical Division of Stryker Corporation

Prime 5th Wheel Stryker Medical Stretchers Model 1105 Intended to transport patients in a horizontal position.

DateJul 22, 2011 (initiated Jul 19, 2011)
StatusTerminated
ReasonThe firm has determined that the brake rod assemblies of various stretchers that were built with a 5th wheel were manufactured with roll pins that did not meet the materials specifications. Some roll pins may have fractured during the assembly process and could fall out of the brake rod assemblies, rendering the brake or steer systems inoperable from certain pedals or causing a reduction in overa
Root causeNonconforming Material/Component
ActionStryker Medical sent an "Urgent Medical Device Correction" dated July 19 ,2011 to all affected customers. The letter included; the name and model numbers of the stretchers, the reason for the correction and dates of manufacture. It contained the instructions to check the braking system and if found to be inoperable, to remove it from service, and if operable to check prior to each use. Also, information regarding contacting the recipients of any loaned or sold stretchers to relay the information in the letter and Medwatch reporting information. A self addressed postcard to confirm receipt
Quantity80
DistributionWorldwide Distribution: Nationwide (USA) including the states of NY, MI, LA and CA ,and the countries of Canada, China, Spain, Japan, Korea, United Kingdom, South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.