ClearanceBrief
Recall Z-2949-2011 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems

BrightView XCT Gamma Camera, 3/4". Philips Medical Systems, Inc. Intended to produce images depicting the anatomical distributions of single photon emitting radioisotopes within the human body for interpretation by medical personnel.

DateAug 4, 2011 (initiated Jun 15, 2011)
StatusTerminated
ReasonNon-conformances with BrightView X/XCT systems running version 2.0.1 including data transfer, XCT acquisition, Pixel Mask for image reconstruction and table moving incorrectly.
Root causeSoftware design
ActionPhilips Medical Systems, Inc. set a "Customer Information" letter dated July 8, 2011 to all affected customers. The letter informed them of an upcoming software updated (version 2.5). For questions please contact 800-722-9377 (option 5).
Quantity213 units
DistributionWorldwide Distribution: (USA) Nationwide distribution and the countries of Greece, Australia, Belgium, China, Czech Republic, France, Germany, Indonesia, Israel, Japan, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Puerto Rico, Saudi Arabia, Slovakia, South Korea, Spain,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080927BRIGHTVIEW VCT IMAGING SYSTEM
Philips Medical Systems (Cleveland), Inc., Apr 11, 2008