Recall Z-2947-2011 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems

BrightView X Gamma Camera, 3/8". Philips Medical Systems, Inc. Intended to produce images depicting the anatomical distributions of single photon emitting radioisotopes within the human body for interpretation by medical personnel.

DateAug 4, 2011 (initiated Jun 15, 2011)
StatusTerminated
ReasonNon-conformances with BrightView X/XCT systems running version 2.0.1 including data transfer, XCT acquisition, Pixel Mask for image reconstruction and table moving incorrectly.
Root causeSoftware design
ActionPhilips Medical Systems, Inc. set a "Customer Information" letter dated July 8, 2011 to all affected customers. The letter informed them of an upcoming software updated (version 2.5). For questions please contact 800-722-9377 (option 5).
Quantity213 units, all systems
DistributionWorldwide Distribution: (USA) Nationwide distribution and the countries of Greece, Australia, Belgium, China, Czech Republic, France, Germany, Indonesia, Israel, Japan, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Puerto Rico, Saudi Arabia, Slovakia, South Korea, Spain,