Recall Z-3106-2011 ยท LXH Orthopedic Manual Surgical Instrument

DePuy Mitek, Inc., a Johnson & Johnson Co.

Depuy Mitek Ratchet Handle with Quick Connect Product code: 219970. Instrument used with anchor inserters for insertion of DePuy Mitek Interference Screws (orthopedic).

DateAug 31, 2011 (initiated Jun 27, 2011)
StatusTerminated
ReasonQuick Disconnect Ratchet Handle may slip up the driver shaft or disengage leaving the shaft in the screw head.
Root causeProcess change control
ActionDepuy Mitek issued Original Customer Notification Letter on 6/27/11 and an Updated letter to clarify the lot number information sent 6/29/11. Hospitals requested to immediately check all inventories for affected product and return for repalcement product. Also notify surgeons aware of this potential issue and potential risk of a delay in surgery. questions or concerns with regard to this program, please contact your local DePuy Mitek Representative or DePuy Mitek Customer Service directly at 1-800-382-4682, PROMPT 3.
Quantity75 units
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.