Recall Z-2715-2011 ยท FPB Filter, Infusion Line

Pall Life Sciences Puerto Rico Llc

Pall Total Nutrient Admixture Filter, 1.2 micron filter This filter is a 1.2 micron low protein binding membrane indicated for the removal of inadvertent particulate debris and microorganisms (Candida Albicans) from total nutrient admixtures and undiluted intravenous fat emulsions. Product is designed for use with intravenous infusion pumps only.

DateJun 30, 2011 (initiated Jun 15, 2011)
StatusTerminated
ReasonIncorrect filter media used (0.2 um filter used instead of the required 1.2 um) within the TNA1 filter used with total parenteral nutrition mixtures.
Root causeLabeling mix-ups
ActionPall Life Sciences LLC, sent an "URGENT MEDICAL DEVICE RECALL" letter dated June 15, 2011 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed customers to maintain the product in quarantine. If distributed, the letter request that customers notify the user to discontinue use. Additionally, customers were asked to complete and return an attached Recall Response Form. The letter also included instructions on how to return the product. Please contact the Customer Service Department at 516-801-9923, if you have
Quantity15,575 units
DistributionWorldwide Distribution--USA (nationwide) including the states of CA, CO, CT, IL, KS, LA, MA, MO, MS, NC, NY, OK, SD, TX, WA, and WY, and Puerto Rico and the countries of Colombia and Costa Rica.