Recall Z-2979-2011 ยท LNH System, Nuclear Magnetic Resonance Imaging

Siemens Medical Solutions USA, Inc

MAGNETOM Aera/Skyra System - Nuclear magnetic resonance imaging

DateAug 10, 2011 (initiated Jun 20, 2011)
StatusTerminated
ReasonFirm discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making servicing necessary.The potential exists for a cable connection to become overheated, leading to the emission of smoke and possible material damage.
Root causeNonconforming Material/Component
ActionSiemens sent a "CUSTOMER SAFETY ADVISORY NOTICE" letter dated June 20, 2011 to all affected customers. The letter describes the product, problem, and actions that will be taken. The letter informs customers that an inspection of the gradient cable connections and an emergency shutdown mechanism will be installed by a field service representative during a visit to all affected locations.
Quantity26
DistributionNationwide Distribution-USA (nationwide) including the states of AK, FL, GA, IL, MA, MI, MN, MT, NC, NJ, NY, OH, TX, UT, VA, WA, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101347MAGNETOM AERA; MAGNETOM SKYRA
Siemens Medical Solutions USA, Inc., Oct 1, 2010