Recall Z-2914-2011 ยท KPS System, Tomography, Computed, Emission
Philips Medical Systems (Cleveland) Inc
Vertical Brake Hubs component of the Philips MX8000 Series Dual v. Exp. Computed Tomography System, Serial Numbers: 8008-8413, 80001-800022, Model #728130, 510 (k) #K010817. Product Usag: The MX 8000 series Dual v. Exp are Computed Tomography X-Ray system units intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles a
| Date | Aug 1, 2011 (initiated Jun 3, 2011) |
| Status | Terminated |
| Reason | Philips Medical Systems decided to conduct a recall/field correction of the patient support vertical brake hub component in all of their CT MX8000 Dual v. Exp Computed Tomography X-Ray System scanners due to a component defect which could cause an unexpected downward motion of the table (couch) while the patient is on the table. This event could occur because the short key connecting the vertical |
| Root cause | Device Design |
| Action | Philips Medical Systems (Cleveland) Inc. issued an "URGENT - Medical Device Correction - Field Safety Notice" letter dated June 1, 2011 to all affected customers, via Certified Mail for tracking purposes. The notification identified the product, the problem involved, and the action(s) to be taken by the customers with regard to this device issue. The notification instructs the customers to monitor their units and if the recalled product malfunctions, to immediately notify their Phillips service representative. The letter also states that a Philips Field Service Engineer will visit each custome |
| Quantity | 412 system units were distributed. |
| Distribution | Worldwide Distribution -- USA (nationwide) including the states of: AR, AZ, CA, CO, FL, FA, HI, IL, IN, KY, MA, MI, MO, MS, NC, NV, NY, OH, OK, SC, SD, TN, TX, VA, VT, WA and Puerto Rico and the country of Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K010817 | MX8000 V5.0 CT SYSTEM Philips Medical Systems (Cleveland), Inc., Jun 15, 2001 |