Recall Z-0131-2012 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Beckman Coulter Inc.
SYNCHRON CX7 Delta Clinical System Part #: 468205 Product Usage: The Beckman SYNCHRON CX4 DELTA, CX7 DELTA, CX4CE, and CX7 systems are fully automated and computer controlled instruments designed for the in vitro diagnostic quantitation of biological fluid components and therapeutic drugs as well as the qualitative determination of drugs of abuse in urine
| Date | Oct 27, 2011 (initiated Aug 23, 2010) |
| Status | Terminated |
| Reason | The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems. |
| Root cause | Software design |
| Action | Beckman Coulter sent an "URGENT: PRODUCT CORRECTIVE ACTION" letter dated September 20, 2010 with an attached Fax Back Response Form (via US Postal Service for US Customers) to all customers who purchased the Synchron CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems. The letter provides the customers wtih an explanation of the problem identified and actions to be taken. Customers were instructed to share this information with laboratory staff and retain this notification as part of their laboratory Quality System documentation. If they forwarded the affected product listed above to an |
| Quantity | 1070 systems total in US |
| Distribution | Worldwide Distribution - USA (nationwide) and the country of Canada |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K950958 | SYNCHON CX DELTA CLIN SYST (CX4, CX5, CX7 DELTA) Beckman Instruments, Inc., May 8, 1995 |