Recall Z-0127-2012 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

SYNCHRON CX5 Clinical System Part #: 759300 Product Usage: The Beckman SYNCHRON CX4 DELTA, CX7 DELTA, CX4CE, and CX7 systems are fully automated and computer controlled instruments designed for the in vitro diagnostic quantitation of biological fluid components and therapeutic drugs as well as the qualitative determination of drugs of abuse in urine

DateOct 27, 2011 (initiated Aug 23, 2010)
StatusTerminated
ReasonThe recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems.
Root causeSoftware design
ActionBeckman Coulter sent an "URGENT: PRODUCT CORRECTIVE ACTION" letter dated September 20, 2010 with an attached Fax Back Response Form (via US Postal Service for US Customers) to all customers who purchased the Synchron CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems. The letter provides the customers wtih an explanation of the problem identified and actions to be taken. Customers were instructed to share this information with laboratory staff and retain this notification as part of their laboratory Quality System documentation. If they forwarded the affected product listed above to an
Quantity1070 systems total in US
DistributionWorldwide Distribution - USA (nationwide) and the country of Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K881495SYNCHRON CX(TM)5 CLINICAL ANALYZER
Beckman Instruments, Inc., Jun 1, 1988