Recall Z-2842-2011 ยท NPD

Ansell Healthcare Products LLC

ENCORE(R) UNDERGLOVE, Latex, Powder-Free, Smooth Surgical Gloves, REF 2018465, Size 6 1/2, Sterile, Manufactured for Ansell Healthcare, 1635 Industrial Road, Dothan, AL 36303, Made in Sri Lanka Surgical gloves

DateJul 18, 2011 (initiated Jun 10, 2011)
StatusTerminated
ReasonGloves do not meet the tensile requirements of ASTM D 3577 reducing the shelf life of the product from 5 years to 2 years.
Root causeOther
ActionAnsell Healthcare Products LLC sent an Urgent Voluntary Recall letter dated June 9, 2011, to all affected customers on June 10 and 13, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disconstinue use and distribution of any of the affected product and return them to the firm. If the customer distributed gloves further, customers were advised to forward a copy of the notification letter along with their tracking number to their customers. Customers were instructed to complete and return the attached Response Form
Quantity176960 pairs
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051793ENCORE STERILE POWDER-FREE LATEX SURGICAL GLOVE WITH PROPRIETARY COATING (CONTAINING GLYCERINE)
Ansell Healthcare Products, LLC, Aug 17, 2005