ClearanceBrief
Recall Z-2860-2011 ยท MEB Pump, Infusion, Elastomeric

I-Flow Corporation

I-Flow ON-Q with Select-A-Flow Pumps, Intended for continuous and/or intermittent delivery of medication to and around surgical wound sites and/or close proximity to nerves for preoperative, perioperative and postoperative regional anesthesia and pain management. It is a controller that allows the user to regulate the amount of medication that is dispensed.

DateJul 28, 2011 (initiated Apr 15, 2011)
StatusTerminated
ReasonThe recall was initiated because I Flow became aware that a small percentage of ON-Q Pump with Select A Flow may develop a minor leak with a flow rate controller. There may be consequences for the patient, such as under delivery of pain medication or the need to prematurely remove the pump, which would result in loss of the pain treatment provided by the pump.
Root causeProcess control
ActionI-Flow Corporation sent an "Important Voluntary Market Withdrawal Notice" dated May 12, 2011 to all affected customers. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to quarantine and return all impacted products within their facilities. Customers were instructed to respond with in 5 days of receiving the Market Withdrawal Notice and fax it to (949) 206-2663. Customers with additional questions, were instructed to call I-Flow Customer service at (800) 448-3569 (8-5pm, Pacific Standard time) or I-Flow 2
Quantity96,892 units total (95,637 units in US)
DistributionNationwide (USA) Distribution and worldwide to: Australia, Canada, Cyprus, Italy, Germany, Finland, Greece, Israel, South Africa, Sweden, UAE, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063530ON-Q, PAINBUSTER, C-BLOC, SELECT-A-FLOW, ONDEMAND, HOMEPUMP, ECLIPSE, C-SERIES, ONE-STEP KVO, EASYPUMP
I-Flow Corp., Jan 26, 2007