Recall Z-2879-2011 ยท HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment

W & H DentalWerk Buermoos GMBH

Arthrex Reciprocating Saw Attachment Type: V-600SR Distributed by: Arthrex Inc., Naples, Florida United States The system Is a modular electric system, consisting of a handpiece drive, various accessories, and a charging station for machining bones. It can be used in the following areas of application: orthopaedic and A&E Interventions, such as osteotomies, large and small bone operations, and j

DateJul 25, 2011 (initiated May 24, 2011)
StatusTerminated
ReasonReciprocating saw attachment V 600 Sr (REF: 17306210), screws can become loose or come off (the attachment) and fall into the surgical wound and may not be noticed due to its small size.
Root causeEquipment maintenance
ActionMadrimed GmbH (Distributor partner) of W&H Dentalmechanik GmbH sent a letter dated May 24, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were ask to advise their users to check the adapters thoroughly before and after each application, until the conversion of all attachments have been completed and to send all products that need to be converted to the nearest service centers. W&H decided to convert all products that might be affected in their service centres and equip them with an additional security device. If you have any questions
Quantity194 units
DistributionWorldwide distribution: USA (nationwide) and countries of: Austia, Australia, Canada, Czech Republic, Denmark, France, Germany, Italy, Kuwait, Malaysia, Netherlands, Russia, Slovakia, South Africa, Spain, Taiwan, Turkey and United Kingdom.