Recall Z-2977-2011 ยท IYE Accelerator, Linear, Medical
Siemens Medical Solutions USA, Inc
ARTISTE/RT Therapist v4.2 and/or Oncologist v4.2 systems. Manufactured by Siemens AG, Medical Solutions, Kemnath, Germany. Delivery of X-ray photon and electron radiation for the therapeutic treatment of cancer.
| Date | Aug 10, 2011 (initiated Jun 3, 2011) |
| Status | Terminated |
| Reason | A potential safety rick related to transferring reference images for offset calculation between a syngo RT Therapist v 4.2 or a syngo RT Oncologist 4.2 and other versions of the COHERENCE/syn go RT Therapist or Oncologist or vice versa. An incorrect offset may be applied for patient positioning and result in dose being delivered to the wrong location. |
| Root cause | Software design |
| Action | Siemens sent an "URGENT: MEDICAL DEVICE CORRECTION/CUSTOMER SAFETY ADVISORY NOTICE" letter dated June 3, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter provides preventive measure instructions for customers to use until an update for the system is available. |
| Quantity | 121 devices |
| Distribution | Worldwide Distribution-USA (nationwide) including the states of CA, FL, NC, NY, OH, PA, SD, UT, and WI and countries of Australia, Belgium, Croatia, Czech Republic, Germany, India, Ireland, Japan, Poland, Saudi Arabia, South Africa, Spain, Sweden, and United Kingdom. |
| FDA record | Open on fda.gov |