Recall Z-2893-2011 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems, Inc. Oncology Systems

Clinac Linear Accelerators; Varian Medical Systems, Palo Alto, CA. Radiation Therapy intended to deliver megavoltage x-ray treatments for conventional radiotherapy and stereotactic radiosurgery and radiotherapy.

DateJul 28, 2011 (initiated May 24, 2011)
StatusTerminated
ReasonThe throat cover on the High Energy Clinac may detach if not properly installed and possibly strike a patient.
Root causeDevice Design
ActionVarian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated May 24, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter provides directions on how to re-install the gantry center throat cover. Customers can contact their Varian Customer Support Manager at 1-888-827-4265 for further clarification.
Quantity278
DistributionWorldwide Distribution-including USA (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100890VARIAN HIGH ENERGY LINEAR ACCELERATOR
Varian Medical Systems, Jul 13, 2010