Recall Z-2888-2011 ยท DWJ System, Thermal Regulating

Cincinnati Sub-Zero Products Inc

WarmAir Model 135 Convective Warming Device, Cincinnati Sub-Zero Products, Inc. Used to prevent hypothermia and/or reduce cold discomfort before, during, and after surgical procedures through convective heat transfer from the controller to a warm-air heated blanket

DateJul 27, 2011 (initiated Jun 1, 2011)
StatusTerminated
ReasonDue to its design the circuit breaker on the device could trip causing it to shut down and not provide heated air to the blanket.
Root causeDevice Design
ActionCincinnati Sub-Zero Products, Inc. sent an 'URGENT Medical Device Recall Field Action Notice" dated July 12, 2011 to all affected customers. The letter includes: description of product, problem, and information to request a repair tubing kit free of charge. It also asks customers to complete and return a "Recall Response Form". For additional information e-mail WA135-2011@cszinc.com or call (800) 989-7373 or (513) 772-8810.
Quantity5474 units
DistributionWorldwide Distribution: (USA) Nationwide, and the countries of Australia, Brazil, Bulgaria, Canada, Chile, Ecuador, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Malaysia, Mexico, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Saudi Arab
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101148WARM AIR MODEL 135 HYPERTHERMIA SYSTEM AND BLANKETS
Cincinnati Sub-Zero Products, Inc., Oct 1, 2010