ClearanceBrief
Recall Z-2793-2011 ยท DQY Catheter, Percutaneous

Abbott Vascular

Guidant RAPIDO Cut-Away Guiding Catheter, 6F. Part Numbers: 66676-201 and 667552-100. For use in cardiac diagnostic and therapeutic procedures.

DateJul 12, 2011 (initiated Jan 7, 2011)
StatusTerminated
ReasonFive lots of Guide Catheters and Accessory products were processed through Ethylene Oxide (ETO) sterilization with additional plastic material covering the pallet, which is not consistent with validated packaging configuration for ETO sterilization.
Root causeProcess control
ActionAbbott Vascular sent an Urgent Device Recall letter, dated January 7, 2011, noting the products being recalled and the reasons for recall. The firm also provided the following reconciliation instructions to its territory manager: -Help customer identify aflected product. - CaIl Abbott Vascular Customer Service (800) 221-9902 to receive RGA number. Record RGA number below. - Fax this completed form to (951) 914-3826. - Return a copy of this completed form with the returned product
Quantity389 units total
DistributionNationwide distribution, and Ireland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021455RAPIDO GUIDING CATHETER, 6F MODEL# 6776; RAPIDO GUIDING CATHETER, 8F MODEL# 6775, 6777, 6778, 6779, 6780, 6781, 6782,
Guidant Corp., Aug 2, 2002
K031459RAPIDO CUT-AWAY CUTTER
Guidant Corp., Jul 23, 2003
K031505RAPIDO CUT-AWAY GUIDING CATHETER
Guidant Corporation, Jun 25, 2003
K031688RAPIDO CUT-A-WAY ROTATING HEMOSTASIS VALVE
Guidant Corporation, Jul 2, 2003
K031903RAPIDO CUT-AWAY BLEEDBACK CONTROL VALVE (BBCV), MODEL 7568
Guidant Corporation, Mar 3, 2004