Recall Z-2792-2011 ยท DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass

Abbott Vascular

Abbott Vascular COPILOT Kit, COPILOT Bleedback Control Valve Part Number : 1003330. For use in cardiac diagnostic or therapeutic procedures.

DateJul 12, 2011 (initiated Jan 7, 2011)
StatusTerminated
ReasonFive lots of Guide Catheters and Accessory products were processed through Ethylene Oxide (ETO) sterilization with additional plastic material covering the pallet, which is not consistent with validated packaging configuration for ETO sterilization.
Root causeProcess control
ActionAbbott Vascular sent an Urgent Device Recall letter, dated January 7, 2011, noting the products being recalled and the reasons for recall. The firm also provided the following reconciliation instructions to its territory manager: -Help customer identify aflected product. - CaIl Abbott Vascular Customer Service (800) 221-9902 to receive RGA number. Record RGA number below. - Fax this completed form to (951) 914-3826. - Return a copy of this completed form with the returned product
Quantity389 units total
DistributionNationwide distribution, and Ireland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K991102COPILOT BLEEDBACK CONTROL VALVE, MODEL 1003331
Guidant Corp., Jun 9, 1999