Recall Z-2766-2011 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems, Inc. Oncology Systems

TrueBeam Linear Accelerators (aka Trilogy Mx) Varian Medical Systems, Palo Alto, CA Product Usage: Intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors , and conditions anywhere in the body where radiation treatment is indicated.

DateJul 8, 2011 (initiated May 19, 2011)
StatusTerminated
ReasonImaging arms of the TrueBeam Accelerator may have loose encoder pulleys that could lead to inaccurate readout of arm geometry.
Root causeDevice Design
ActionVarian Medical Systems, Inc. Oncology Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letters on May 19, 2011, via certified mail, return receipt requested in the US. The letter identified the product, problem and recommended actions to be taken. For additional information contact Varian Customer Support at 1-888- 827-4265.
Quantity77
DistributionWorldwide Distribution -- USA (nationwide) and countries of: Canada, Denmark, France, Germany, Isreal, Italy, Netherlands, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092871TRILOGY MX
Varian Medical Systems, Inc., Nov 30, 2009