ClearanceBrief
Recall Z-2808-2011 ยท FRN Pump, Infusion

Zevex International, Inc.

***EXPANDED RECALL*** Curlin Infusion Administration Set, REF 340-4175, Sterile EO, Zevex, Inc., 4314 Zevex Park Lane, Salt Lake City, UT 84123. The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.

DateFeb 25, 2011 (initiated Jan 24, 2011)
StatusTerminated
ReasonDefect in tubing may trigger false Air-In-Line warning.
Root causeProcess control
ActionThe firm, MOOG, sent an "Expansion of Moog/Curlin Ambulatory Pump Administration Set Recall" letter dated May 2, 2011 to its customers. This letter informed the customers that this recall has been expanded to included specific lot numbers of Curlin model ambulatory infusion administration sets. The customers were instructed to follow the outlined process in the letter: remove suspect product code from inventory, contact MOOG Customer Service at 1-800-970-2337, prompt #7, contact your distributor to arrange replacement product, request a shipping return label as per the distributor's process a
Quantity2,020
DistributionWorldwide Distribution: USA including states of: AR, CA, FL, IA, KS, MS, NE, NY, OH, OK, TN, TX, and UT; and countries including: Australia, Croatia, France, Germany, Kuwait, New Zealand, Qatar, Saudi Arabia, South Africa, United Arab Emirates, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981816CUE AMBULATORY VOLUMETRIC INFUSION PUMP
Becton Dickinson Curlin, LLC, Dec 4, 1998