ClearanceBrief
Recall Z-2764-2011 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems, Inc. Oncology Systems

Clinac Linear Accelerator; Model numbers H14, H27 and H29. Product Usage: Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

DateJul 8, 2011 (initiated May 18, 2011)
StatusTerminated
ReasonThe Coolant System input water supply manifold may leak in some Clinac Linear Accelerators, posing a risk of electrical shock to any person working within the protective housing.
Root causeOther
ActionVarian Medical Systems, Inc. Oncology Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letters on May 18, 2011, via certified mail, return receipt requested. The letter identified the product, problem and recommended actions to be taken. The letter states that customers will be contacted by Varian Service to arrange for replacement of the recalled product. For additional information contact Varian Medical Systems, Inc at 888- 827-4265.
Quantity93
DistributionWorldwide Distribution -- USA (nationwide) and countries of: Australia, Columbia, Denmark, France, Ireland, Italy, Japan, Morocco, Poland, Portugal, Saudi Arabia, South Korea, Spain, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100890VARIAN HIGH ENERGY LINEAR ACCELERATOR
Varian Medical Systems, Jul 13, 2010