Recall Z-2693-2011 ยท MSD Catheter, Hemodialysis, Implanted

Medical Components, Inc dba MedComp

THD155428- 15.5F X 28cm Titan HD(TM) Catheter Kit Product Usage: Hemodialysis catheter insertion kit

DateJun 28, 2011 (initiated May 19, 2011)
StatusTerminated
ReasonProduct was packaged with the incorrect valved peelable introducer. The catheter sized is 15.5F and the introducer packaged in the kit is a 15F. The kit should contain a 16F introducer.
Root causeMixed-up of materials/components
ActionMedcomp sent a "PRODUCT ALERT" letter dated May 19, 2011 to all affected customers. The letter described the product, problem, and actions to be taken. The letter is requesting the return of all affected product. Customers are instructed to contact their customer service representative for an RGA (returned good authorization) number. If you have any questions or concerns contact Medcomp at 215-256-4201 ext. 2225.
Quantity90
DistributionWorldwide Distribution -- USA (nationwide) including states of AZ, CA, CO and RI and countries of Australia, Brazil, and Netherlands