Recall Z-2418-2011 ยท DQO Catheter, Intravascular, Diagnostic
Boston Scientific Corporation
Boston Scientific iCross, 40 MHz Coronary Imaging Catheter, 3.6F (1.18mm) x 135 cm, REF / Catalog No. 51805, UPN / Product No. H749518050, Sterilized using irradiation, Made in USA 47900 Bayside Parkway, Fremont, CA 94538 This catheter is intended for ultrasound examination of coronary intravascular pathology only. Intravascular Ultrasound imaging is indicated in patients who are candidates for
| Date | Jun 10, 2011 (initiated Mar 28, 2011) |
| Status | Terminated |
| Reason | Catheter tip detachment that is not associated with elongation of the catheter typically caused by excessive pulling forces. EXPANDED: Boston Scientific is initiating a recall of all iCross Coronary Imaging Catheters because they have received and confirmed 8 occurrences of catheter tip detachments due to embrittlement. |
| Root cause | Process design |
| Action | Boston Scientific " Important Customer Notification" letters dated March 14, 2011 were hand delivered to customers by Boston Scientific sales representatives starting on March 28th, 2011. The sales representatives verbally reviewed information. A tracking log will be maintained document completion of the customer notifications. The letter described the product, problem and possible root cause as well as one corrective action, a change in sterilization method from Gamma to E-beam implementation. Boston Scientific sent an Urgent Medical Device Recall Removal - Immediate Action Required letter |
| Quantity | 66,033 |
| Distribution | Worldwide Distribution - USA (nationwide) and Puerto Rico and the countries of Japan, Trinidad, Tobago, and American Virgin Islands. |