Recall Z-2744-2011 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Cambridge Endoscopic Devices, Incorporated

Cambridge Endoscopic Autonomy Laparo-Angle 5mm Metzenbaum Scissors with Monopolar Cautery, 45 cm Catalog Number: LA1101-12 Intended for use in a variety of minimally invasive procedures to facilitate grasping, mobilization, dissection, and transection of tissue.

DateJul 6, 2011 (initiated May 20, 2011)
StatusTerminated
ReasonPET shrink sleeve insulation material dislodging from the distal tips of Autonomy Laparo-Angle (ALA) Metzenbaum Scissors
Root causeProcess control
ActionCambridge Endoscopic Devices, Inc. sent an "IMPORTANT MEDICAL DEVICE RECALL" letter dated May 20, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Customers are instructed to return unused inventory using RGA Number 301. Returned devices will be replaced. Questions regarding this recall, call 508-405-0790 for assistance.
Quantity255 units
DistributionWorldwide Distribution--USA (nationwide) and the countries of Armenia , Australia, Austria, Chile, China, Denmark, Hong Kong, India, Ireland , Italy, Japan, Saudi Arabia, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061425PUREWRIST SCISSORS, MODEL PW1101-01;PUREWRIST DISSECTOR, MODEL PW1201-01;PUREWRIST HOOK, MODEL PW1301-01