Recall Z-2980-2011 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA, Inc.

Symbia Radiation Therapy Pallet. Siemens Medical Solutions USA, Inc. An accessory to the Symbia T Series Emission Computed Tomography System/X-Ray Computed Tomography System.This specialty pallet is used for performing SPECT and/or SPECT/CT Scans on oncology patients with the goal of fusing the SPECT and/or SPECT/CT data with CT simulation data for planning radiation therapy. The pallet attache

DateAug 10, 2011 (initiated Jun 22, 2011)
StatusTerminated
ReasonThere a potential patient safety issue when using the Symbia Radiation Therapy Pallet. When a patient is not iso-centered before starting a scan with the Symbia T Series Systems and a non-circular orbit is selected, it is possible that one of the detectors may come into contact with the patient.
Root causeDevice Design
ActionSiemens Medical Solutions USA, Inc. sent an "Urgent Field Correction Recall" letter dated June 06, 2011 to all affected customers. The letter informs customers of the potential safety issue if the patient is not iso-centered before starting a scan and a non-circular orbit is selected and provided recommendations including ensure the patient is iso-centered before starting a scan. If the patient cannot be positioned at the iso-center of the RTP, instructions were provided. The customers are asked to complete the Mandatory Fax-Back Form acknowledging receipt and understanding of the letter.
Quantity18 pallets
DistributionWorldwide Distribution: Nationwide (USA) including the states of Florida and Texas; and the countries of Canada, China, Italy, Mexico, Poland, Portugal and Russia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082506SYMBIA, VERSION 4.0
Siemens Medical Solutions USA, Inc., Nov 21, 2008