Recall Z-2615-2011 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Terumo Cardiovascular Systems Corporation
Cardiovascular Procedure Kit Part number 70667 Eisenhower Army Med Center Heart Lung Pack, Sterile, Terumo Cardiovascular Systems Ashland, MA. The pressure monitoring kit is indicated for sterile single use with the reusable pressure transducer.
| Date | Jun 22, 2011 (initiated May 3, 2011) |
| Status | Terminated |
| Reason | The firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's manufacturing process. |
| Root cause | Other |
| Action | Terumo Cardiovascular Systems sent an URGENT MEDICAL DEVICE REMOVAL notice dated May 3, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. Customers were sent 10 replacement pressure monitoring kits with the removal notice as immediate replacements. The firm will send additional replacements to customers that have more than 10 units of affected product as noted on the Customers' returned reponse form. Customers that received the custom packs were provided stickers to identify the recalled packs. The custome |
| Quantity | 1515 ALL CUSTOM PACKS |
| Distribution | Worldwide Distribution -- USA (nationwide )including the states AR, CA, FL, GA, MO, MT, NC, ND, OK, OR, PA, TN, TX, UT, VA, WA and WY, and the countries of Argentina, Belgium, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatamala, Honduras, Hong Kong, Indonesia, Kuwait, Malaysia, Mexi |
| FDA record | Open on fda.gov |