Recall Z-2787-2011 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Siemens Medical Solutions USA, Inc

Siemens YSIO - solid state x-ray imager Solid state x-ray imager (flat panel/digital imager), image-intensified fluoroscopic x-ray system

DateJul 11, 2011 (initiated Apr 28, 2011)
StatusTerminated
ReasonThere is a potential hazard to staff or patients when lowering the table height on the Ysio system. If staff or patients lower extremities are placed underneath the table and the user lowers the table top, it is possible that parts of the table may collide with extremities and cause injuries.
Root causeOther
ActionSiemens sent a Safety Advisory Notice to all affected customers. The Safety Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to place the Safety Notice in the System's instruction for use. If product was distributed further, customers were to forward the safety notice to the new owner and inform the firm of the new owner. Customers were instructed to complete the Acknowledgement of Receipt Form and return to the firm.
Quantity141
DistributionNationwide Distribution including Washington, DC and the following States: AR, CA, CO, FL, ID, IL, IN, IA, KS, KY, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NY, NC, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081722YSIO
Siemens Medical Solutions USA, Inc., Aug 22, 2008