Recall Z-2570-2011 ยท NBW System, Test, Blood Glucose, Over The Counter

Roche Diagnostics Operations, Inc.

ACCU-CHECK PERFORMA STRIP NO. 05987288031, 100 BANDELETTES REACTIVES, ROCHE DIAGNOSTICS GMBH, WWW.ACCUCHEK.COM glucose monitoring test strips.

DateJun 21, 2011 (initiated May 18, 2011)
StatusTerminated
ReasonErroneous results: Internal investigations revealed that the test strips exhibited a performance issue that can cause test strip errors or a bias in test result that exceed the tolerance range according to EN ISO 15197.
Root causeComponent design/selection
ActionDirect accounts were notified by a Roche Diagnostics press release on 5/18/2011 that included the product's name, lot number, and information to withdraw the remaining inventory of lot 470049 from storage
Quantity27,019 units
DistributionNo US distribution, in France Only.