Recall Z-2786-2011 ยท LNH System, Nuclear Magnetic Resonance Imaging

Siemens Medical Solutions USA, Inc

Siemens MAGNETOM Systems Symphony, Sonata and Symphony a TIM System Nuclear magnetic resonance imaging system

DateJul 11, 2011 (initiated Mar 31, 2011)
StatusTerminated
ReasonFirm has discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation.
Root causeNonconforming Material/Component
ActionSiemens initated their recall of the device on March 31, 2001, by sending a Customer Safety Advisory Notice/Letter to affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were notified that detectors will be installed in appropriate locations on the system, which will switch off the system's power electronics in the event of a malfunction in the power supply of the gradient coils, preventing damage to the system and its environment.
Quantity440
DistributionNationwide Distribution including AL, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, and WI and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993731SONATA GRADIENTS/MAGNETOM SYMPHONY SYSTEM
Siemens Medical Solutions USA, Inc., Dec 23, 1999