Recall Z-2665-2011 ยท DRF Catheter, Electrode Recording, Or Probe, Electrode Recording

Biosense Webster, Inc.

HALO Catheter with Auto 10 Technology, Part Number D-1160-43-S, Catalog Number D7T20282CT The affected catheters are indicated for multiple electrode electrophysiological mapping of cardiac structures of the heart, i.e. recording or stimulation only.

DateJun 27, 2011 (initiated May 23, 2011)
StatusTerminated
ReasonThe recall was initiated because Biosense Webster, Inc. has recently become aware that 20 electrodes catheters with Auto ID Technology are not being recognized when connected to the CARTO 3 System.
Root causePending
ActionBiosense Webster, Inc. sent a Customer Notification Letter, dated May 24, 2011, with an Acknowledgement Form and Instructions for Manually Entering the Catheter Identification in the CARTO 3 System to all US and Outside US (OUS) customers who purchased these products. For customers in the US, two letters and forms will be sent via express mail, addressed to the EP/Cardiology Lab and Risk Management Office, respectively. Instructions for manual entry of the catheter identification in the CARTO 3 System are being included in the communication to customers. These instructions will allow custome
Quantity431units (298 in US)
DistributionWorldwide Distribution - USA (nationwide) and the countries of Belgium, Japan, and Singapore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953768CORDIS WEBSTER A20 DIAGNOSTIC DEFLECTABLE TIP CATHETER
Cordis Webster, Inc., Nov 22, 1995