Recall Z-2626-2011 ยท MDB System, Blood Culturing
Becton Dickinson & Co.
BACTEC" FX System Software, catalog #441398 and #441750 The BD BACTEC" FX instrument is designed for the rapid detection of bacteria and fungi in clinical specimens. Samples are drawn from patients and injected directly into BACTEC culture vials, which are placed into the instrument for incubation and testing.
| Date | Jun 23, 2011 (initiated Aug 10, 2010) |
| Status | Terminated |
| Reason | In vitro diagnostic instrumentation may exhibit random operational errors that could delay effective patient treatment. |
| Root cause | Software design |
| Action | BD Diagnostics Systems sent an "URGENT PRODUCT RECALL" letter dated August 2010 to all affected customers. The letter describes the product, problem and action to be taken. The letter described the type of error messages the customers may experience. All affected units will be scheduled for on site software update as well as a disk drive replacement. A BD field service engineer will contact the customers to schedule the on site visit. Customers experiencing issues prior to the scheduled visit or have questions should contact the BD Diagnostics Service Department at 800-638-8663. |
| Quantity | 907 units |
| Distribution | Worldwide Distribution--USA (nationwide) including countries of Australia, Belgium, Canada, China, Colombia, New Zealand, Hong Kong, Japan, Korea, Philippines, Singapore, Taiwan, and Thailand. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.