Recall Z-2591-2011 ยท OUB Instrument For Treatment Of Hyperhidrosis

Miramar Labs, Inc.

MiraDry System consisting of the DTS3000 Console, MiraDry Handpiece (MD3001-HP), the miraDry biotip and the miraDry Treatment Template system. Miramar Labs, Inc. 445 Indio Way Sunnyvale, CA 94085 Intended use: Treatment of primary axillary hyperhidrosis.

DateJun 17, 2011 (initiated Apr 24, 2011)
StatusTerminated
ReasonReports of skin burns on axilla after treatment with the miraDry System.
Root causeDevice Design
ActionThe firm,Miramar, has a service provider in Japan that assists in dealing with customers. The service providers were provided a copy of the investigation report and communicated directly with the customers to set up appointments for the field correction. The customers were personally visited by a representative from Miramars service provider in Japan. All devices had field correction performed on April 26-28, 2011.
Quantity8 units
DistributionInternational only: Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103014MIRADRY SYSTEM
Miramar Labs, Inc., Jan 28, 2011