Recall Z-2585-2011 ยท LFL Instrument, Ultrasonic Surgical

Integra Limited

Integra CUSA NXT Ultrasonic Tissue Ablation System Service Module The CUSA NXT Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation. emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable. including Neurosurgery. Gastrointestinal and affiliated organ surgery. Urological surgery. Plastic and Reconstructive surgery. General sur

DateJun 17, 2011 (initiated May 2, 2011)
StatusTerminated
ReasonThe insulated covering of a wire cable that powers the vacuum pump in the CUSA NXT Service Module can become abraded over time resulting in the exposure of bare wire. If this bare wire were to come in contact with the metal surface of the pump, it could result in a short circuit and a loss of aspiration.
Root causeDevice Design
ActionIntegra sent a "FIELD ACTION NOTIFICATION" dated May 2, 2011 to all affected customers. The letter identified the product, the problem and the actions to be taken. Customers were informed of a Field Corrective Action that is being initiated. All devices (customer owned, Integra Loaner Pool, and Sales demos) will undergo rework of the vacuum pump assembly. The rework will be scheduled, at customer sites by Integra Field Service Technicians or returned to a designated Integra Service Center for rework. A loaner will be presented to customer accounts upon request. The letter asks that each
Quantity105 units
DistributionWorldwide -- USA (nationwide) and countries of France, Germany, Hungary, Ireland, Italy, Netherlands, Spain, Sweden and Taiwan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081459CUSA SELECTOR NXT ULTRASONIC TISSUE ABLATION SYSTEM
Integra Radionics, Inc., Aug 13, 2008