Recall Z-2795-2011 ยท MSB System And Accessories, Isolated Heart, Transport And Preservation

Vitrolife, Inc.

XVIVO Perfusion AB (Part of Vitrolife), THAM Solution, 3.3 mmol/ml, 50 mL, REF 99019, Sterile, Used to increase the pH of the Perfadex used for perfusion of donor lungs. Perfadex with THAM kits, 8 x 1000mL REF 19017, and 2 x 2800mL REF 19018.

DateJul 12, 2011 (initiated May 12, 2011)
StatusTerminated
ReasonPrecipitation observed in some bottles of Tham buffer used for adjusting pH of Perfadex prior to perfusion of donor lungs.
Root causeProcess control
ActionXVIVO Perfusion, part of the Vitrolife Group, sent out a "Partial recall of Perfadex with THAM" letter on May 13, 2011 to all affected customers. Customers were instructed to to quarantine all Perfadex with THAM. Replacement THAM was to be shipped by Vitrilife within a few days and the exchanged THAM was to immediately sent to Vitrolife, Englewood, CO. For additional Information please call (303) 762-1933
Quantity1457 bottles
DistributionWorldwide distribution including USA and Europe.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081997PERFADEX WITH THAM
Vitrolife Sweden AB, Oct 9, 2008