Recall Z-2725-2011 ยท DSI Detector And Alarm, Arrhythmia

Brentwood Medical Technology Corp

IQecg IQdevices SDK, Part Numbers: 4-100-1403, Software Version 8.3.2 Product Usage: IQecg is to be used, under the supervision of a Physician, to obtain Electrocardiograms from the human body surface.

DateJul 14, 2011 (initiated May 18, 2010)
StatusTerminated
ReasonThe recall was initiated because when a user of the IQmanager, IQdevices SDK, and IQiC (Centricity) software versions listed above accesses the QT dialogue box to either accept or edit the QT value, the QTc value is recalculated using a Hodges QTc equation, regardless of patient's age. The software does this without notifying the user of the change from use of the Bazett equation, which is utiliz
Root causeSoftware design
ActionThe firm, Midmark, sent an "URGENT: MEDICAL DEVICE FIELD CORRECTIVE ACTION" letter on May 19, 2010 with an attached Corrective Return Material Authorization (RMA) form via US Postal Service Certified Mail to its consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately examine their IQecg, IQmanager, or lQiC Centricity software to determine the software version currently in use. If it is a listed product, it is subject to this field corrective action. A sample software label was attached to aid the consignees/customers
Quantity8 units
DistributionWorldwide distribution: USA (Nationwide) and countries of: Algeria, Australia, Canada, France and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K955023BRENTWOOD PC-ECG
Brentwood Medical Products, Inc., Jan 22, 1997