Recall Z-2740-2011 ยท OOI Real Time Nucleic Acid Amplification System

bioMerieux, Inc.

NucliSens EasyQ Enterovirus serotype 68 v1.1: Catalog number: 200300 In Vitro Diagnostic Test

DateJul 6, 2011 (initiated May 10, 2011)
StatusTerminated
ReasonThe NucliSENS EasyQ Enterovirus Reagents are unable to detect Enterovirus serotype 68, in respiratory samples.
Root causeDevice Design
ActionBioMerieux Vitek, Inc. sent an "Urgent Product Correction Notice" letter on/about May 10, 2011. The letter included; a description of the product, problem, possible impact to the patient and recommendations. Recommendations included instructing customers to discontinue the use of the kit for the detection of Enterovirus 68 and Rhinovirus 87 serotypes and noted the corrected IFU will be available the first week of April 2011. An Acknowledgement form was attached to be returned via fax. For additional questions please contact your local BioMerieux Clinical Customer representative at (800) 68
Quantity42 kits
Distribution(USA) Nationwide distribution including the states of MD, MI, MN, MT, NY and PA