ClearanceBrief
Recall Z-2758-2011 ยท KOG

Ethicon, Inc.

J-Vac Bulb Suction Reservoir, Manufactured for: Ethicon, Inc. Somerville, New Jersey, 08876-0151. Closed Wound Drainage Systems have been used as an adjunct in the following specialties to evacuate potentially detrimental collection of certain fluids (e.g. pus, extravascular blood, bile) from wounds in body cavities and to reduce the risk of infection.

DateJul 7, 2011 (initiated Mar 24, 2011)
StatusTerminated
ReasonThere is a potential for the sterile barrier to be compromised in multiple lots of Blake silicone drains, Blake silicone drain kits, Blake cardio connectors, J-VAC reservoirs, J-VAC drain adaptors.
Root causePackaging process control
ActionEthicon, Inc. sent an "URGENT VOLUNTARY PRODUCT RECALL" letter dated March 24& 25, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to discontinue use of the products and to return the products to Stericycle. The letter includes shipping labels for product return. Ethicon will send replacement product. Customers are asked to return a completed Business Reply Card to Stericycle regardless of whether or not they have the products subject to this recall. Customers with questions regarding t
Quantity846,990
DistributionWorldwide Distribution-- USA (nationwide) and Puerto Rico and the countries of Armenia, Australia, Austria, Azerbaijan, Columbia, Cyprus, Czech Republic, Denmark, French Guinea, French Polynesian, Germany, Israel, Italy, Japan, Kuwait , Netherlands, New Caledonia, New Zealand, Norway, Singapore, Slo
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953655J-VAC SUCTION RESERVOIRS & J-VAC BLAKE SILICONE DRAINS
Johnson & Johnson Medical, Inc., Oct 13, 1995